modified name
|
Modify code | operate |
Online Oligo Analyzer Description:
1. Accuracy
The properties of oligonucleotide chains (such as Tm values and extinction coefficients) are derived from fitting formulas using relevant algorithms and are for reference only, not representing actual experimental data. Different tools may use different formulas and parameters, so the results may vary slightly. If degenerate bases are present in the sequence, the calculation is based on the average value of the base data represented by the degenerate bases.
2. COMPLEMENT
Displays the inverse complementary sequence of the input sequence, in the 5-3 direction. Terminal modification groups are automatically ignored for inverse complementary sequences; however, if the modification group is a basic base analog, it is calculated as a basic base. The complementary base of deoxyxanthine (dI) is cytosine (C).
3. LENGTH
The number of single nucleotide residues in the oligonucleotide chain, including if the modifying group is a single nucleotide residue analogue.
4. GC CONTENT
The molar percentage of guanine and cytosine nucleotides in the oligonucleotide chain.
5. TM
Melting temperature (Tm) refers to the temperature at which two complementary base sequences are 50% melted. In the Sangon Bioprimer Analyzer, Tm calculation is carried out according to the nearest neighbor two-state model, and the calculation formula (1) adopted is:
Where ΔH and ΔS refer to the enthalpy and entropy of this oligonucleotide calculated based on thermodynamic melting parameters, respectively. The thermodynamic parameters adopted by the bioengineering primer analyzer are from the article published by Allai and Santa Lucia in Biochemistry in 1997 (DOI: 10.1021/bi962590c); R is the ideal gas constant (1.987 cal/K·mole); [Oligo] refers to the effective molar concentration of the oligonucleotide. We assume that the oligonucleotide chains are not self-complementary, and the effective molar concentration is taken as 1/4 of the total oligonucleotide concentration. Here, the default total oligonucleotide concentration is 500 nM.
Salt ion correction: Monovalent salt ions will have a certain influence on the Tm value of the primer. The salt concentration condition in formula (1) is 1M Na ions, so the obtained Tm(K) is the thermodynamic temperature (unit: K) at a sodium ion concentration of 1M. However, in most actual cases, such a high concentration is not possible, so it is necessary to correct the concentration. The salt ion correction formula used by the bioengineering primer analyzer comes from the article published by Owczarzy in Biochemistry in 2004 (DOI: 10.1021/bi034621r). Formula (2) is:
Where Tm* is the ion-corrected Tm value, fGC is the GC content of the oligonucleotide chain, and [Na+] is the actual ion concentration. The default Na ion concentration in the bioengineering primer analyzer is 50 mM. After obtaining Tm*, converting the unit to degrees Celsius gives the final predicted Tm value.
6. MOLECULAR WEIGHT
Molecular weight refers to the mass of 1 mole of oligonucleotide chain and is calculated as the accumulation of the masses of all atoms in the oligonucleotide chain.
7. EXTINCTION COEFFICIENT
The molar extinction coefficient (ε260) refers to the absorbance of 1 mole of oligonucleotide chain in aqueous solution at 260 nm. This value determines the relationship between the oligonucleotide chain mass and the number of ODs synthesized. When the base composition is random, it can be approximated that 1 OD of DNA double strands is 50 μg, and 1 OD of single-stranded oligonucleotides is 33 μg. However, for customer-customized oligonucleotide chains, calculations need to be performed based on the sequence. The bioengineering primer analyzer uses a nearest-neighbor two-state model to calculate the extinction coefficient, referencing Michael's 2003 article published in *Nucleic Acids Research* (DOI: 10.1093/nar/gnh015).
8. nmole/OD260 and µg/OD260
Based on the molar extinction coefficient, the molecular weight and mass of the oligonucleotide chain contained in each OD can be calculated. The calculation formulas are: nmole/OD260=106/ε260 and µg/OD260 = nmole/OD260 × molecular weight × 10-3.